Fluke Corporation MET/CAL Procedure ============================================================================= INSTRUMENT: Fluke 30: (1 year) CAL VER /5500 DATE: 02-May-97 AUTHOR: Fluke Corporation REVISION: $Revision: 1.3 $ ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 11 NUMBER OF LINES: 88 CONFIGURATION: Fluke 5500A ============================================================================= # # Source: # Fluke Model 30 Clamp Meter Service Information (PN 929042 January 1994). # # Compatibility: # 5500/CAL 4.0 or MET/CAL 4.0 or later # # Subprocedures: # None # # Required Files: # 55_30.bmp # # System Specifications: # TUR calculation is based on specification interval of the accuracy file. # The default 5500A accuracy file contains 90 day specs. # # Fluke makes no warranty, expressed or implied, as to the fitness # or suitability of this procedure in the customer's application. # # The 90 day specifications of the 5500A are used in TUR computations. # STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 DISP {System accuracy of 5500A + actual 50-turn} 1.001 DISP {toroid used must be entered for each ACC statement} 1.002 DISP {to reflect actual system accuracy. } 1.003 DISP {See procedure README file.} 1.004 ASK- P F W 1.005 ASK+ K 1.006 HEAD EQUIPMENT SETUP 1.007 DISP This procedure requires a 50-turn Toroid Coil. 1.007 DISP Refer to the Model 30 Clamp Meter Service Information 1.007 DISP for directions for constructing a suitable coil. 1.008 HEAD {CURRENT} 1.009 PIC 55_30 1.010 HEAD CURRENT: {200A Range} 1.011 5500 20.0A 60H SI S T5 # System accuracy of 5500A + actual 50-turn toroid used must be entered here. 1.012 ACC 20.A 1.2U 1.013 MEMI Enter UUT reading in Amps: 1.014 MEME 1.015 MEMC 200 A 1.2U 60H 2.001 5500 190.0A 60H SI S T5 # System accuracy of 5500A + actual 50-turn toroid used must be entered here. 2.002 ACC 190.0A 2.7U 2.003 MEMI Enter UUT reading in Amps: 2.004 MEME 2.005 MEMC 200 A 2.7U 60H 3.001 HEAD CURRENT: {400A Range} 3.002 5500 40A 60H SI S T5 # System accuracy of 5500A + actual 50-turn toroid used must be entered here. 3.003 ACC 40A 4U 3.004 MEMI Enter UUT reading in Amps: 3.005 MEME 3.006 MEMC 400 A 4U 60H 4.001 5500 350A 50H SI S T5 # System accuracy of 5500A + actual 50-turn toroid used must be entered here. 4.002 ACC 40A 8U 4.003 MEMI Enter UUT reading in Amps: 4.004 MEME 4.005 MEMC 400 A 8U 50H 5.001 5500 400A 50H SI S T5 # System accuracy of 5500A + actual 50-turn toroid must be entered here. 5.002 ACC 40A 12U 5.003 MEMI Enter UUT reading in Amps: 5.004 MEME 5.005 MEMC 400 A 12U 50H 6.001 HEAD {VOLTAGE} 6.002 HEAD VOLTAGE: {200V Range} 6.003 5500 200 20.0V 0.5U 60H SI 2W 7.001 5500 200 190.0V 2.6U 60H SI 2W 8.001 HEAD VOLTAGE: {600V Range} 8.002 5500 600 60V 4U 60H SI 2W 9.001 5500 600 600V 10U 60H SI 2W 10.001 HEAD {RESISTANCE} 10.002 5500 200 190Z 3.1U 2W 11.001 END